Analysis of Residual Solvents in Pharmaceuticals by Water-Soluble Samples Using H2 Carrier (USP 467)


User Benefits

  • The use of H2 as a carrier gas, which is inexpensive and easy to obtain, contributes to cost reduction in the laboratory.
  • H2 carrier gas is used to achieve the accuracy required by the USP.
  • A hydrogen sensor ensures H2 carrier gas can be used safely.
  • Using a state-of-the-art HS-20 NX headspace sampler enables reliable measurement of residual solvents in pharmaceuticals.

Introduction

Test methods for residual solvents in pharmaceuticals are specified in the USP (United States Pharmacopeia) General Chapter <467> Residual Solvents, which mainly specifies using the headspace GC method. Residual solvents in pharmaceuticals are classified into Classes 1 to 3 based on potential risks to human health and strictly controlled accordingly. Sensitive analysis is required. The carrier gas generally used for analysis is He, but the depletion of He has become a problem recently. Consequently, there is a need for using an alternative carrier gas such as H2 for analysis. Any method changes, such as substituting He with an alternate carrier gas, must be validated according to USP Chapter <1467> Residual Solvents—Verification of Compendial Procedures and Validation of Alternative Procedures.

This paper presents the results of using H2 as a carrier gas with an HS-20 NX headspace sampler to analyze Class 1 and 2 watersoluble samples in accordance with USP General Chapter <467> Residual Solvents.

*Analytical samples were prepared using a different procedure fromthe USP to confirmthe performance of the apparatus.

Instrument Configuration and Analysis Conditions

A Nexis GC -2030 gas chromatograph and Shimadzu HS-20 NX headspace gas sampler were connected to measure the listed Class 1 and Class 2 standard solutions according to USP General Chapter <467> Residual Solvents, Procedure A. The analytical conditions for GC and HS are indicated in Table 1.

Table 1 Water-Soluble Sample Analysis Conditions

GC Analysis Conditions (Procedure A)

Model: Nexis GC-2030
Detector: FID-2030 flame ionization detector
Column: SH-Rxi™-624 Sil MS (0.32 mm I.D.×30 m, d.f.= 1.8 μm)
Column Temperature: 40 °C (20 min) – 10 °C/min – 240 °C (20 min) Total 60 min
Injection Mode: Spilt 1:5
Carrier Gas Controller: Constant linear velocity mode (H2 and He)
Linear Velocity: 35 cm/sec
Detector Temperature: 
250 °C
FID H2 Flowrate: 
32 mL/min
FID Make-up Flowrate: 
24 mL/min (N2)
FID Air Flowrate:
200 mL/min

HS Analysis Conditions (Procedure A)

Oven Temperature: 80 °C
Sample Line Temperature: 110 °C
Transfer Line Temperature: 120 °C
Vial Stirring: Off
Vial Volume: 20 mL
Vial Heat-Retention Time: 45 min
Vial Pressurization Time: 1 min
Vial Pressure: 75.0 kPa(N2)
Loading Time: 0.5 min
Needle Flush Time: 5 min
Injection Volume: 1 mL
Load Equilib. Time: 0 min

Analysis of Class 1 Standard Solution (Water-Soluble Sample)

The analysis results for Procedure A using a H2 carrier are shown in Fig. 1. The S/N ratio and repeatability values for each peak are shown in Table 2. Table 3 shows the S/N ratio and repeatability values for each peak when using a He carrier as a reference.

Using a H2 carrier, the results obtained with Procedure A satisfied the requirements of the USP, which specifies that "the S/N ratio for 1,1,1-trichloroethane in the Class 1 standard solution is not less than 5."

Fig. 1 Chromatogram of Class 1 Standard Solution by Procedure A Water-Soluble Sample.png
Fig. 1 Chromatogram of Class 1 Standard Solution by Procedure A (Water-Soluble Sample)
Peak Compound S/N Ratio*1(n=4) %RSD*1(n=4)
1 1,1-Dichloroethane 93 1.99
2 1,1,1-Trichloroethane 123 1.45
3 Carbon tetrachloride 15 3.94
4 Benzene 145 2.48
5 1,2-Dichloroethane 57 2.60

Table 2 S/N Ratio and Repeatability of Class 1 Standard Solution (Procedure A) Using H2 Carrier

Peak Compound S/N Ratio*1(n=4) %RSD*1(n=4)
1 1,1-Dichloroethane 131 1.44
2 1,1,1-Trichloroethane 150 1.64
3 Carbon tetrachloride 13 7.30
4 Benzene 188 0.66
5 1,2-Dichloroethane 76 0.81

Table 3 S/N Ratio and Repeatability of Class 1 Standard Solution (Procedure A) Using He Carrier
*1 The S/N ratio and relative standard deviation (%RSD) values are reference values and not intended to be guaranteed values.

Analysis of Class 2 Standard Solution (Water-Soluble Sample)

Fig. 2 shows the analysis results for a Class 2A and Fig. 3 for a Class 2B standard solution. Methyl isobutyl ketone (MiBK), a component newly added to USP General Chapter <467> Residual Solvents for Class 2A was also measured. (Results with the H2 carrier are indicated in black, the He carrier in pink, and for MiBK with the H2 carrier in blue.) Satisfactory system suitability results were also obtained, for which the USP specifies that “the resolution between acetonitrile and methylene chloride in the Class 2A standard solution is not less than 1.0“ when using Procedure A.

*The resolution values shown in Fig. 2 are intended as reference values and are not guaranteed.

Fig. 2 Chromatogram of Class 2A Standard Solution by Procedure A (Water-Soluble Sample).png
Fig. 2 Chromatogram of Class 2A Standard Solution by Procedure A (Water-Soluble Sample)
Fig. 3 Chromatogram of Class 2B Standard Solution by Procedure A Water-Soluble Sample.png
Fig. 3 Chromatogram of Class 2B Standard Solution by Procedure A (Water-Soluble Sample)

Conclusion

Using a H2 carrier gas, the analysis achieved the accuracy levels required by USP Chapters <467> and <1467>. The Nexis GC- 2030 features a hydrogen sensor to ensure safe H2 usage. Hydrogen sensors detect potential leaks early. When hydrogen leakage increases, the main power supply is switched off to prevent accidents. Using a H2 carrier can help reduce lab costs. For information about using a N2 carrier for analysis of residual solvents in pharmaceuticals using water-soluble samples, refer to Application News No. G325.

Related Solutions

Nexis GC-2030

Witness the capabilities of our Living Technology in Nexis GC-2030! Designed with the analyst in mind, Nexis GC-2030 provides phenomenal performance beyond your expectation.

nSMOL-01.jpg
HS-20 NX Headspace Sampler

With the state-of-the-art HS-20 NX headspace sampler, rest assured to obtain reliable measurement of residual solvents in pharmaceuticals.

Resources You May Like
nSMOL-01.jpg
nSMOL Antibody BA Kit

nSMOL (nano-surface and molecular orientation limited proteolysis) is Shimadzu’s proprietary, innovative technique that enables selective proteolysis of the Fab region of monoclonal antibodies.

IRTracer-100_01.jpg
IRTracer-100

This system achieves excellent sensitivity with an SN ratio of 60,000:1, high resolution at 0.25 cm-1, and high-speed scanning capable of 20 spectra/second. The performance of medium and higher end models is supported by high reliability including advanced dynamic alignment and an interferometer with a dehumidifier.

  • As a trusted partner in the pharmaceutical industry, Shimadzu provides a variety of instruments and suitable workflows for drug impurity analysis to help ensure compliance to these guidelines.
  • This handbook shares a wide variety of methodologies for different classes of drug impurities using different equipment like LC-MS/MS, UHPLC, GC-MS, GC-MS/MS, ICP-MS
  • These methods are in sync with relevant ICH guideline and provides a simple solution for scientists who seek confident analyses.

What's in it?
-Genotoxic Impurities
-Organic Impurities
-Elemental Impurities
-Residual Solvent
-Impurities from Packaging

 
White Paper
  • As a trusted partner in the pharmaceutical industry, Shimadzu provides a variety of instruments and suitable workflows for drug impurity analysis to help ensure compliance to these guidelines.